コンテンツメニュー

Quantitative chemical analyses of rocks by X-ray fluorescence spectrometry using a mixed alkali flux (Li_2B_4O_7 : LiBO_2=7 : 3)

Center of Instrumental Analysis, Yamaguchi University Volume 8 Page 16-23
published_at 2000
KJ00000199469.pdf
[fulltext] 664 KB
Title
<研究報告>混合融剤(A12)を使用した岩石の蛍光X線分析
Quantitative chemical analyses of rocks by X-ray fluorescence spectrometry using a mixed alkali flux (Li_2B_4O_7 : LiBO_2=7 : 3)
Creators Umemoto Kengo
Creators Imaoka Teruyoshi.
Creators Yamasaki Teruyoshi.
Creators Owada Masaaki.
Source Identifiers
A systematic XRF analytical routine for major and trace elements of silicate rocks using a mixed alkali flux (Li_2B_4O_7 : LiBO_2= 7 : 3,Spectromelt A12) has been made with the Rigaku RIX3000 introduced in the Center of Instrumental Analysis, Yamaguchi University. With usage of the Spectromelt A12 (Merck), a homogeneous glass bead could be made even in an infusible quartz-rich rock. In order to make calibration curves, we used 32 geochemical standard samples issued by Geological Survey of Japan, U.S. Geological Survey, Korean Institute of Energy and Resources and S. Africa Bureau of Standards, and the analytical results were a fairly good in comparison with the recommended values for a wide variety of rocks ranging from siliceous to ultramafic ones. Repeated analyses of JB-1a and JG-1a using the calibration curves proved that the analytical system is very well also in the reproducibility. Using this method (named UME5), major and trace element analyses were possible using one single glass bead.
Languages jpn
Resource Type departmental bulletin paper
Publishers 山口大学機器分析センター
Date Issued 2000
File Version Version of Record
Access Rights open access
Relations
[NCID]AN1046804X
Schools 機器分析センター