コンテンツメニュー

Development of a palpation sensor for detecting prostatic cancer and hypertrophy : optimum structural design of sensor

Transactions of the Japan Society of Mechanical Engineers. C Volume 64 Issue 618 Page 571-576
published_at 1998-02
2010010175.pdf
[fulltext] 476 KB
Title
前立腺癌および前立腺肥大症触診用センサの開発研究 : センサの構造設計
Development of a palpation sensor for detecting prostatic cancer and hypertrophy : optimum structural design of sensor
Creators Chonan Seiji
Creators Jiang Zhong Wei
Creators Tanahashi Yoshikatu
Creators Tanaka Mami
Creators Kato Takuji
Creator Keywords
palpation sensor PVDF film measurement of organic stiffness prostatic cancer and hypertrophy detection
This paper is concerned with the development of a palpation sensor for detecting the prostatic cancer and hypertrophy. The sensor is a layered medium with two polyvinylidene fluoride (PVDF) films embedded in the matrix rubber block. The temporal mean values, I_1 and I_2, of absolute voltage signals from the polymer films are calculated and their ratio I_1/I_2 is used as the sensor output. First, the fundamental characteristics of sensor are examined on a rectangular sensor with a plane surface. Then, the surface geometry of sensor is designed so that the sensor output clearly extracts the features on the stiffness of the materials. Finally, a compact cylindrical sensor which is inset in the doctor's forefinger is presented and its characteristics as a palpation sensor are studied experimentally. Some representative clinical data on the hardened and normal prostate glands are also presented.
Languages jpn
Resource Type journal article
Publishers 日本機械学会
Date Issued 1998-02
Rights
Copyright: 日本機械学会 (The Japan Society of Mechanical Engineers)()
File Version Version of Record
Access Rights open access
Relations
[ISSN]0387-5024
[NCID]AN00187463
[isVersionOf] [URI]http://www.jsme.or.jp/publish/ronbun/
Schools 大学院理工学研究科(工学)